How to Use the iSurfa-220 Surface Roughness Tester: Practical tips for surface roughness measurement
Scan QR

Surface roughness decides whether a seal holds oil, a bearing runs quiet, or a coating bonds — and no caliper will ever measure it. That is the job of a surface roughness tester. Here is the full operating routine for the iSurfa-220, from charging to a certified calibration check.

1. Know what the number means
The instrument evaluates the profile against the parameter definitions in ISO 4287 and ISO 4288, the framework most drawings worldwide use; North American prints often cite ASME B46.1.

  • Ra — arithmetic mean profile deviation in µm or µin. The most common drawing callout because it is stable and easy to compare.
  • Rz — maximum profile height within a sampling length. Reacts to deep scratches and tool marks, so it suits sealing faces and hydraulic spools.
  • Rq — root mean square height, still written as RMS on older US drawings.

Two surfaces can share the same Ra and still behave completely differently, so the parameter and the cut-off length must match the drawing before any reading is trustworthy.

2. Set up the instrument
Charge the battery fully and let the tester reach the same temperature as the workpiece; drift from a cold unit looks exactly like poor repeatability. Power on, wait for the self-test to finish without pressing keys, then set the parameters: cut-off length (λc), five sampling lengths for the evaluation length, the parameter set, µm or µin, and the Gaussian filter.
For turned, milled, and ground surfaces, ISO 4288 ties the cut-off to the expected roughness. Most machined parts run λc 0.8 mm over a 4 mm evaluation length; finer surfaces use 0.25 mm, rougher or cast surfaces 2.5 mm and above. The wrong cut-off length can move Ra by twenty percent or more on the same surface, so record it on the inspection sheet.

iSurfa 220 Surface RoughnessTester 图(1) iSurfa 220 Surface RoughnessTester 图(2) 

3. Fit and position the diamond stylus
Seat the diamond stylus — typically 5 µm tip radius on a 90° cone — until the locating pin engages, then lock it finger-tight. Handle it by the shank, never the tip. A stylus that looks seated but is slightly loose scatters readings and can break on the first traverse.
Clean the surface with a lint-free cloth and isopropyl alcohol, rest the skid flat, and traverse across the lay rather than along the tool marks. Clamp the part, keep the bench free of running machines, and do not rest your hand on the instrument — a few micrometres of vibration show up in the profile.

iSurfa 220 Surface RoughnessTester 图(4) iSurfa 220 Surface RoughnessTester 图(5) iSurfa 220 Surface RoughnessTester 图(6)

4. Measure and record
Press measure and keep everything still until the display settles. Take three readings at different positions and work with the average; one traverse is a spot check, not a surface roughness measurement. Transfer the stored records to a PC through the data interface and keep the cut-off with every value — that is what makes an inspection record auditable.

5. Calibrate and maintain
Surface roughness tester calibration is a habit, not an annual formality. Check against a certified specimen close to your working range at the start of each shift or at least weekly, and send the unit for an accredited calibration once a year. Afterwards, power down, cap the stylus, wipe the housing, and store the instrument dry in its case.

Bottom line
Match the parameter and cut-off length to the drawing, keep the stylus clean and seated, control the setup, and verify against a specimen. Do that, and the iSurfa-220 delivers repeatable numbers your customers will accept without argument.
If a callout or probe selection is unclear, send the print to the Mikrosize application team — we will match the configuration to the specification you actually have to meet.

Leave a Comment

Your email address will not be published. Required fields are marked *

×
WeChat QR

Wechat: 13918745376