Dear Customer,
Thank you for choosing the iLeeb-150 Pen-Type Leeb Hardness Tester. This precision-engineered instrument will give you accurate, convenient hardness testing on a wide range of materials. This guide explains how to use it, step by step.
1. Preliminary checks
Before you start, carry out the necessary preliminary checks. First, inspect the instrument’s appearance. Pick up the hardness tester and check the housing for damage, scratches or deformation. This is not only a matter of appearance: the integrity of the housing directly affects the safety of the precision components inside. If you find any problem, do not use the instrument. Contact our customer service team and we will resolve the issue for you.
Next, check the display. The screen is the main interface for reading measurement data, so it must be clear and free of defects. Look for dead pixels, bright spots, or a dim, blurred display. A faulty screen can make you misread a result and draw the wrong conclusion from your work.
Then check the impact device. It is a core component of the hardness tester and directly determines measurement accuracy. Examine the impact ball: it should be smooth and show no sign of wear. A worn ball, pitted or deformed, is like a blunt cutting tool — the readings will not be accurate. At the same time, make sure the impact body moves freely in the guide tube without sticking. Push it gently to confirm that the movement is smooth. If anything is abnormal, replace the part promptly.
The battery level also deserves attention. Check the battery indicator on the screen. If the charge is low, recharge the instrument with the original charger. A flat battery in the middle of an important measurement is a needless interruption, so keep the battery charged before every job.
2. Accessories and calibration
Once the preliminary checks are complete, select the right accessories and calibrate the instrument. Match the support ring to the material: whether the shape is regular or irregular, whether the part is thick or thin, and how rough the surface is. For a rough surface, use a support ring with a large contact area to keep the measurement stable. For thin-walled parts, use a light ring that fits closely, so that the material is not damaged.
-1.png)
-1.png)
Then install the correct impact device. The iLeeb-150 is normally supplied with several impact devices, such as the D-type and the DC-type. Each type suits a particular hardness range and type of material. When installing, make sure the device is firmly seated. Shake it gently to check for looseness: a loose impact device will shift during measurement and the readings will be inaccurate.
Calibration is essential to accuracy. Before routine measurement, calibrate the tester with a standard hardness block whose hardness is close to that of the material you are going to test. Hold the instrument vertical and press the impact device firmly against the block. Start a measurement and take at least three readings. Record each value, calculate the average, and compare it with the certified value of the block. If the deviation exceeds the permitted tolerance (typically ±2%), enter calibration mode and adjust the instrument as described in the operation manual. After calibration, take a check measurement to confirm that the instrument is reading correctly.
-8.jpg)
-5.jpg)
3. Measurement
With everything ready, you can begin measuring. Press the power key and wait for the screen to light up and finish initializing. During initialization, the instrument automatically checks its internal circuits, sensors and other components. When the normal measurement screen appears, you can proceed.
Select the measurement mode that suits the material and the task. The usual options are single-point measurement and continuous measurement. Single-point mode is enough when you need the hardness at one point; continuous mode is more efficient when you need readings at several points in quick succession.
Next, select the hardness scale. The iLeeb-150 measures in Leeb hardness (HL) and supports conversion to Brinell (HB), Rockwell (HRA, HRB, HRC) and Vickers (HV). Base your choice on the material and the relevant industry standard, so that the scale, and the result, is correct.
-1-300x300.png)
With the scale set, press the impact device against the test surface and press the release button to trigger the impact. Read the value on the screen, and record it together with the material, the scale and the test position, so that the result can be traced later.