Operating Procedure for the iFlaw-80 Ultrasonic Flaw Detector
Scan QR

Dear users, before operating the Ultrasonic Flaw Detector, please follow the procedure below to ensure accurate inspection results and stable operation of this ultrasonic flaw detector machine supplied by Mikrosize. The ultrasonic flaw detector uses described here apply to weld, forging, casting and plate inspection in both production and in-service testing.

I. Preparation before Flaw Detection

  •  Equipment Inspection

Before power-on, perform a full equipment inspection. Use a magnifying glass to examine the housing for deformation, damage or cracks to guarantee protective performance. Check the display screen for dead pixels and confirm clear, stable display; the numbered ultrasonic flaw detector diagram in the manual identifies every indicator and connector. Test interfaces with a multimeter to verify sound electrical connections and stable signal transmission, and ensure probe connectors are tight with good contact. Stop operation immediately and contact qualified service personnel if any abnormality occurs.

  •  Probe Selection and Installation

Select probes based on test-piece material, geometry, dimension and expected defect types. A high-frequency probe is preferred for thin-plate testing for higher resolution; a low-frequency probe is selected for thick-walled components to guarantee ultrasonic penetration. Clean the probe and connector to remove contaminants and achieve tight connection. Secure the probe with a holder after installation and manually verify stability.

iFlaw 80 Ultrasonic Flaw Detector图3 4

iFlaw 80 Ultrasonic Flaw Detector图6 4

  • Parameter Setting

Configure flaw-detection frequency, gain, sound velocity and detection depth range on the interface per test-piece features and inspection standards. Understanding the ultrasonic flaw detector working principle — pulse-echo transit time and amplitude evaluation — makes these settings faster and more reliable; a full explanation is available in our ultrasonic flaw detector working principle pdf. Strictly reference standard specifications and double-check every value to secure reliable detection results.

  • Sensitivity Calibration

Before scanning, carry out ultrasonic flaw detector calibration on a reference block of matched material and thickness: plot the DAC or AVG curve as required, set the reference gain and confirm that echoes from the reference reflectors reach the specified height. Recalibrate whenever the probe, material or thickness changes.

iFlaw 80 Ultrasonic Flaw Detector图7

II. Flaw Detection Process

  • Test-piece Surface Preparation

Clean the test-piece surface to remove oil, rust, dust and other contaminants. Grind rough surfaces until surface roughness meets inspection requirements. Apply proper couplant for sufficient probe-to-workpiece coupling and lower detection errors.

  • Scanning Operation

Place the assembled detector steadily against the test-piece surface for tight coupling. Power on the unit and scan with the probe moving at constant speed within defined areas. Closely observe on-screen data and waveforms to identify internal defects and evaluate defect nature and position from waveform features. Mark locations right away once abnormal waveforms appear.

  • Data Recording and Analysis

Record critical data and waveforms in real time during testing. Upon completion, analyze the data against inspection standards and practical experience, judge the acceptance status of the test piece, and assess defect severity and its impact on component performance.

iFlaw 80 Ultrasonic Flaw Detector图8

III. Post-inspection Handling

  • Equipment Cleaning and Maintenance

After testing, wipe the detector and probe using approved cleaning agents and soft cloths to clear surface contaminants. Check all components for looseness or damage; retighten or repair components as needed. Conduct periodic comprehensive maintenance, including parameter calibration and battery care, to maintain good equipment status.

  • Inspection Report Generation

Compile standardized inspection reports from test data and analysis results. Reports shall contain test-piece information, inspection time, methods, parameters, results and suggested countermeasures, and be accurate, clear and complete for follow-up reference.

Following the steps above enables optimal performance of the iFlaw-80 Ultrasonic Flaw Detector. For current ultrasonic flaw detector price, including ultrasonic flaw detector price in india,Malaysia, Russia, Thailand, etc. or an ultrasonic flaw detector for sale enquiry, contact Mikrosize for a quotation and demo video.

Leave a Comment

Your email address will not be published. Required fields are marked *

×
WeChat QR

Wechat: 13918745376